The IceWorm: an improved low-cost, low-power sensor for measuring dissolved CH4 in water bodies

Publikation: KonferencebidragKonferenceabstrakt til konferenceForskningfagfællebedømt

Dokumenter

  • Fulltext

    Forlagets udgivne version, 290 KB, PDF-dokument

Recent studies show emissions of dissolved methane (CH4) in the meltwater from the Greenland Ice Sheet. To better understand the phenomenon and evaluate its potential significance for the Arctic CH4 budget, continuous long-term measurements of dissolved CH4 concentrations are needed. Commercially available dissolved CH4 analyzers (DGEU-UGGA (LGR), CONTROS HydroC CH4 (Kongsberg) and Mini CH4 (pro-Oceanus)) generally have high power consumption and are very costly, limiting their operation in remote off-grid locations.

Here we present calibrations and field tests of a low-cost, low-power alternative – the "IceWorm" - for long-term monitoring of dissolved CH4. The IceWorm uses a Figaro TGS2611-E00 metal oxide sensor (MOS). While MOS are cheap and power efficient, a known drawback is the sensitivity of the sensor's resistance to changes in humidity and temperature. In a previous prototype, we showed that by encasing the MOS in a hydrophobic and gas-permeable silicone membrane, a constant humidity in the headspace around the sensor can be achieved, yielding consistent results when deployed in glacial meltwater at constant temperature (0.0 – 0.1˚C)1. In this updated version, the sensor was encased in a hydrophobic and gas-permeable Teflon membrane allowing for fast (~1 min) equilibrium between the water and headspace around the sensor and hence a rapid detection of changes in dissolved CH4 concentrations.

The first calibration was performed by exposing the IceWorm to stepwise increasing Two field calibrations of the sensor performance in meltwater at 0.0˚C were done: Afterwards, the sensors remained in the field for several weeks in the subglacial meltwater stream and the sensors were recalibrated in lab air under the same conditions to check for long-term sensor drift. Initially, field calibrated to measure dissolved CH4 in glacial meltwater at 0.0˚C, the IceWorm was also tested in a freshwater surface stream at temperatures between 1.6 – 15.7˚C. To account for the temperature difference, we compared the laboratory and field calibrations allowing us to correct the sensor output to temperature variations in the stream.

We will present time series of long-term measurements of dissolved CH4 in two different types of water bodies and discuss the promising performance of the sensor at temperatures different to stable 0˚C as well as the usability of in-air calibrations compared to the field calibrations with discrete samples.

1. Sapper et al. (2022) DOI:10.5194/egusphere-egu22-9972
OriginalsprogEngelsk
Publikationsdato2023
Antal sider2
DOI
StatusUdgivet - 2023
BegivenhedEGU General Assembly 2023: Vienna, Austria & Online - Vienna, Østrig
Varighed: 24 apr. 202328 apr. 2023

Konference

KonferenceEGU General Assembly 2023
LandØstrig
ByVienna
Periode24/04/202328/04/2023

Antal downloads er baseret på statistik fra Google Scholar og www.ku.dk


Ingen data tilgængelig

ID: 336957607